Z-0863-2015: Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrad

Recall record Z-0863-2015 for Bausch & Lomb Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0863-2015
Company in FDA record
Bausch & Lomb Inc
FDA classification date
Dec 23, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

Source facts

Action
Bausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839.
Initiated
2014-11-04
Root cause
Software Manufacturing/Software Deployment
Terminated
2017-03-23
Distribution
Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.
Recall number
Z-0863-2015
Recall status
Terminated
CDRH record ID
131560
FDA classified
2014-12-23
Recall event number
69717

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗