The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
Source facts
Action
The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
Initiated
2005-02-09
Root cause
Other
Terminated
2005-06-29
Distribution
Nationwide and Canada.
Recall number
Z-0864-05
Recall status
Terminated
CDRH record ID
37435
FDA classified
2005-06-02
Recall event number
31232