Z-0864-2019: TRINIAS for diagnostic imaging and interventional procedures in cardiac angiography, neurovascular angiog

Recall record Z-0864-2019 for Shimadzu Medical Systems Usa Com. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0864-2019
Company in FDA record
Shimadzu Medical Systems Usa Com
FDA record creation date
Feb 21, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two issues: Event 1: Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure. Event 2: In rare cases a synchronization error may occur in the data transmission circuit in the device due to external noise. The visibility of the displayed image may be corrupted (images become bit-shifted abnormal images). If this event occurs, it may become difficult to see the object, which may hinder the examination and treatment.

Source facts

Action
The dealers by email on 01/14/2019 and the end users by mail on 01/15/2019. The letters to the end users explain the problem and a workaround. A service representative will contact the end user to schedule a software update.
Initiated
2019-01-14
Root cause
Software design
Terminated
2021-04-30
Distribution
FL, MT, CT, OH, IL, TX, MS, LA, SC
Recall number
Z-0864-2019
Recall status
Terminated
CDRH record ID
170548
FDA record created
2019-02-21
Recall event number
82002

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗