Z-0864-2020: AQURE basic sytem, FLEXLINK, REF 933-599, Version: 2.3.3 (17), CE IVD, Rx Only, UDI: (01)05700699335999

Recall record Z-0864-2020 for Radiometer Medical ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0864-2020
Company in FDA record
Radiometer Medical ApS
FDA record creation date
Jan 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware that there is a potential problem relating to the blood gas and immunoassay analyzer Systems that may result in patient mix-up when connected to some third-party devices. The error may lead to serious adverse health consequences for the patient caused by patient data mix-up.

Source facts

Action
On 07/15/2019, the firm, Radiometer America, Inc., sent an "Urgent: Medical Device Correction" to customers via FedEx with tracking and email to inform them that the firm has become aware that there is a potential problem related to its system that may result in patient mix-up when connected to some third-party devices. This recall affects all system below version 2.3.5. On September 24th, the firm sent a updated letter via email which updated that only 7 third party devices could be potentially affected (Originally 10 third party devices). The firm is instructing customers to: If the customer's affected system is connected to a third-party device, the Recalling Firm is requesting the customer to inform operators to cease performing queries as described in the scenario above until the affected system has been corrected by the Recalling Firm's representative. Additionally, the firm is asking that the customer please complete the Recall Response Form (last page) and return it to the fax number (714-582-0806) or email address provided (RECALL@READIOMETERAMERICA.COM). For any additional questions, customers are instructed to contact their Recalling Firms representative; or 440-925-34875 or email: jgibbons@radioamerica.com.
Initiated
2019-08-08
Root cause
Software design
Terminated
2022-06-06
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV; and countries of: Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, India, Italy, Japan (Tokyo-not a country but it was listed), Netherlands, Norway, Poland, Singapore, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Recall number
Z-0864-2020
Recall status
Terminated
CDRH record ID
178420
FDA record created
2020-01-24
Recall event number
84496

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗