Z-0864-2021: All EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EP

Recall record Z-0864-2021 for Philips Ultrasound Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0864-2021
Company in FDA record
Philips Ultrasound Inc
FDA record creation date
Jan 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The manufacturer has determined that with certain uncommon workflows there is potential for incorrect patient data to be displayed and saved into an exam. The issue(s) manifest differently for different versions of software.

Source facts

Action
On 10/27/2020,the manufacturer/ recalling firm started sending an Urgent Medical Device Correction letter (MDC) with recommended workflows to prevent the potential issues. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction letter. The effectiveness check will include the delivery confirmation of the Urgent Medical Device Correction Letter in the USA or the Field Safety Notice (FSN) worldwide. In addition, effectiveness checks will be conducted by confirmation that the customer has received the letter and have read/understood the actions documented within (customer reply form).
Initiated
2020-10-27
Root cause
Software design
Terminated
2023-04-19
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Cook Islands, Costa Rica, Corte D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Virgin Islands (British), Virgin Islands (U.S.), Yemen, Zimbabwe.
Recall number
Z-0864-2021
Recall status
Terminated
CDRH record ID
184984
FDA record created
2021-01-14
Recall event number
86992

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗