The acoustic display of Mechanical Index and Thermal Index is inaccurate - two transducers may exceed limits for acoustic intensity and MI.
Source facts
Action
On May 6, 2005 the firm mailed a letter, via return receipt requested, to customers. The letter advised of the problem, and provided instructions to create new low power presets. The letter also advised that software would be issued on/before November 2005.
Initiated
2005-05-06
Root cause
Other
Terminated
2008-05-21
Distribution
The units were distributed worldwide. Domestic distribution included hospitals and medical centers.
Recall number
Z-0865-05
Recall status
Terminated
CDRH record ID
39065
FDA classified
2005-06-03
Recall event number
32064