Z-0865-2019: Raindrop Near Vision Inlay, REF: RD1-1, Sterile, Rx Only, Manufacturer ReVision Optics, Inc.

Recall record Z-0865-2019 for RVO 2.0, INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0865-2019
Company in FDA record
RVO 2.0, INC
FDA classification date
Mar 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

During a post approval study the firm found there was an increased risk of corneal haze following the implantation of this device.

Source facts

Action
On 11/13/2018, the firm started emailing Urgent Medical Device Recall notices and acknowledgment forms to consignees. This recall notice was intended for 1) patients implanted with the Raindrop Inlay and 2) eye care providers. Eye care providers were asked to cease further implantation and to return unused affected devices to the recalling firm. Eye care providers were informed that patients implanted with affected devices should continue to see their eye care providers on a regular basis, preferably every 6-months and no less often than every 12 months. Also, patients who have had their devices explanted should maintain a follow-up schedule with their provider. The firm commits to physicians to support the patient's best interests and to continue the FDA post approval study to monitor the device's ongoing performance and safety. The firm requests that all consignees complete and return, acknowledgement forms, regardless of whether they do or do not have unused affected product to return. Acknowledgement forms should be returned via email to Jennifer.Luong@opticsmedical.com. Customers with questions are encouraged to call 949-330-6511.
Initiated
2018-11-13
Root cause
Unknown/Undetermined by firm
Terminated
2024-03-14
Distribution
U.S.: NV, GA, NY, MN, TX, ND, AZ, VA, SD, IN, MA, MO, CA, NC, UT, IL, NH, OH, OK, FL, NJ, TN, MI, PA, HI, ID, KY, NE, AL, LA, CO, MD, MS, OR, WI, AR, WA; No OUS (Foreign) Distribution.
Recall number
Z-0865-2019
Recall status
Terminated
CDRH record ID
169984
FDA classified
2019-03-05
Recall event number
81812

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗