Z-0865-2024: GE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid f

Recall record Z-0865-2024 for GE HEALTHCARE AUSTRIA GMBH & CO. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0865-2024
Company in FDA record
GE HEALTHCARE AUSTRIA GMBH & CO
FDA classification date
Jan 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of an issue with certain IC9-RS ultrasound probes (see Affected Products List) that can result in a double image artifact creating a ghost image with realistic features. An unrecognized artifact may lead to misdiagnosis.

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/22/2023 via letter delivered using traceable means. The notice explained the issue with the device, safety issue, and requested the following actions be taken: 1. Ensure all potential users in your facility are made aware of this correction notification and the recommended actions. 2. You can continue to use your ultrasound system with all other probes. 3. Upon receipt of the letter, verify if your IC9-RS probe is functioning correctly prior to use by performing the Double Image artifact test provided in the letter. (Testing must be repeated monthly) 4. If a double image artifact is seen, do not use the probe and contact a GE HealthCare representative to get a replacement probe. 5. Complete and return the attached acknowledgement form to Recall.79072@ge.com. Please retain this document for your records.
Initiated
2023-12-22
Root cause
Device Design
Distribution
worldwide
Recall number
Z-0865-2024
Recall status
Open, Classified
CDRH record ID
205320
FDA classified
2024-01-30
Recall event number
93727

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗