Negatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.
Source facts
Action
Letters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.
Initiated
2005-04-18
Root cause
Other
Terminated
2005-10-13
Distribution
37 U.S. consignees nationwide.
Ortho-Clinical foreign affiliates in England, France and Spain.
Recall number
Z-0866-05
Recall status
Terminated
CDRH record ID
38786
FDA classified
2005-06-04
Recall event number
31942