Skip to content

Public FDA record

Z-0866-2008: Roche/Hitachi Modular Analytics System, Modular D2400 Module GMMI clinical chemistry analyzer; Catalog nu

Recall record Z-0866-2008 for Roche Diagnostics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0866-2008
Company in FDA record
Roche Diagnostics Corp.
FDA classification date
Aug 13, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Possible mismatch between patient and result.

Source facts

Action
Consignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.
Initiated
2008-03-21
Root cause
Process change control
Terminated
2009-11-18
Distribution
Nationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.
Recall number
Z-0866-2008
Recall status
Terminated
CDRH record ID
69367
FDA classified
2008-08-13
Recall event number
47371

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗