Skip to content

Public FDA record

Z-0866-2012: Gas cylinders used in Gunnell manual wheelchairs with dynamic recline feature, including: Bariatric Rehab

Recall record Z-0866-2012 for Gunnell, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0866-2012
Company in FDA record
Gunnell, Inc.
FDA classification date
Jan 25, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Gunnell has received complaints that the gas cylinders used in Gunnell manual wheelchairs would not hold a firm position, or as described by users, the back support or canes were spongy in the Rehab Tough and Tilt (TNT), Bariatric Rehab Tough and Tilt (BTNT), Rehab Recline and Mobility (RAM), Bariatric Rehab Recline and Mobility (BRAM), and Rehab Kidster.

Source facts

Action
Gunnell, Inc. sent a Medical Device Correction letters dated November 1, 2011 to all consignees. The letter identified the product, problem, and actions to be taken. The letter states that Gunnell Inc will provide free replacement cylinders and reimburse for each correction verified. The notice instructs customers to complete and return the Response Form. Customers were instructed to notify their customers by including a copy of the correction notice. Question and concerns should be addressed to andy.evancho@gunnell-inc.com. or call 1-800-551-0055, attention Andy Evancho.
Initiated
2011-11-01
Root cause
Component design/selection
Terminated
2013-08-29
Distribution
Worldwide Distribution - USA (nationwide) nad the countries of: Australia and Canada
Recall number
Z-0866-2012
Recall status
Terminated
CDRH record ID
105240
FDA classified
2012-01-25
Recall event number
60180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗