Open original FDA record ↗Recall record
- FDA identifier
- Z-0866-2012
- Company in FDA record
- Gunnell, Inc.
- FDA classification date
- Jan 25, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Gunnell has received complaints that the gas cylinders used in Gunnell manual wheelchairs would not hold a firm position, or as described by users, the back support or canes were spongy in the Rehab Tough and Tilt (TNT), Bariatric Rehab Tough and Tilt (BTNT), Rehab Recline and Mobility (RAM), Bariatric Rehab Recline and Mobility (BRAM), and Rehab Kidster.
Source facts
- Action
- Gunnell, Inc. sent a Medical Device Correction letters dated November 1, 2011 to all consignees. The letter identified the product, problem, and actions to be taken. The letter states that Gunnell Inc will provide free replacement cylinders and reimburse for each correction verified. The notice instructs customers to complete and return the Response Form. Customers were instructed to notify their customers by including a copy of the correction notice. Question and concerns should be addressed to andy.evancho@gunnell-inc.com. or call 1-800-551-0055, attention Andy Evancho.
- Initiated
- 2011-11-01
- Root cause
- Component design/selection
- Terminated
- 2013-08-29
- Distribution
- Worldwide Distribution - USA (nationwide) nad the countries of: Australia and Canada
- Recall number
- Z-0866-2012
- Recall status
- Terminated
- CDRH record ID
- 105240
- FDA classified
- 2012-01-25
- Recall event number
- 60180