Z-0866-2020: RayStation stand-alone software treatment planning system. Models 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1,

Recall record Z-0866-2020 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0866-2020
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA classification date
Jan 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD, when intended as source-to-surface distance, it sometimes gives source-to-skin distance.To the best of our knowledge, these issues have not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect dose calculations during treatment planning.

Source facts

Action
Field Safety Notice, Medical Device Correction #52889 #52672 #52971 dated 12/18/2019. ACTIONS TO BE TAKEN BY THE USER " i. Do not use the options Map ROI(s) or Map ROI(s) reversed, only available by right-clicking in the ROI list. Instead, use the Copy ROI geometries dialog available in the Structure Definition module or the Map ROIs dialog available in the Deformable Registration module. " ii. Be aware that the Elekta guard leaf setting from RayStation may be overridden by the R&V system or TCS. Contact the suppliers of your R&V system and TCS for more information. Scripting can be used to position the guard leaves according to the settings of the delivery system to avoid the overrides. " iii. Do not use the beam.GetSSD() scripting method. If needed in scripting, see RayCommunity. " iv. Do not use the SSD value displayed in the DRR in the plan report or exported DRR. Instead, use the SSD value displayed in the Beam data section of the plan report. Please educate planning staff and all users about these workarounds. Inspect your product and identify all installed units with the above software version number(s), then confirm you have read and understood this notice by replying to the notification email. SOLUTION i. The Map ROI issue will be resolved in a future version of RayStation, scheduled for market release in May 2020 (subject to market clearance in some markets). ii. The Elekta guard leaves issue will be resolved in a future version of RayStation, scheduled for market release in May 2020 (subject to market clearance in some markets). iii. The SSD issue will be resolved in the next version of RayStation, scheduled for market release in December 2019 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
Initiated
2019-12-19
Root cause
Software design
Terminated
2020-11-24
Distribution
US Nationwide distribution in states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, WI
Recall number
Z-0866-2020
Recall status
Terminated
CDRH record ID
179082
FDA classified
2020-01-27
Recall event number
84705

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗