Z-0867-2020: Healon GV PRO 0.85 mL, Model (U.S.): 10-2400-14, Model (O.U.S.): 10-2705-14, 10-2715-14, 10-2725-14, and

Recall record Z-0867-2020 for Johnson & Johnson Surgical Vision Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0867-2020
Company in FDA record
Johnson & Johnson Surgical Vision Inc
FDA classification date
Jan 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring additional intervention. Also, reports of clogging of phacoemulsion equipment tubing, which may lead to delay in the procedure or ocular injury.

Source facts

Action
Customers were asked to immediately take the following actions: 1. Compare your inventory against the affected product list. 2. Discontinue using and remove from your inventory all affected lots. 3. Even if you have no inventory, please complete the Customer Reply Form. If you have inventory of any of the affected OVDs products, please complete the Customer Reply Form, noting the lot numbers of the OVDs, and use the shipping label provided to return the product. Returned product will NOT require refrigerated shipping. This notice should be shared with anyone who needs to be aware within your organization or to any organization where the potentially affected products have been transferred. Customers may call 1-877-266-4543 Option 2 if they have additional questions or want to report complaints or adverse events.
Initiated
2019-12-03
Root cause
Labeling False and Misleading
Terminated
2021-04-28
Distribution
U.S.: CA, PA, NJ, GA, IL, MO, CO, TX, FL, AK, NY, HI, UT, LA, OH, IN, MA, NC, AZ, MI, NH, AL, TN, WA, MD, OR, VA, MS, MN, KY, AR, NV, WI, WV, SC, DE, ME, CT, ID, NM, VT, IA, OK, MT, NE, DC. OUS: DE, GB, DK, IT, NL, TR, AT, CH, FR, GR, ES, PT, FO, BE, GP, ZA, IE, HR, HU, IS, NO, SE, FI
Recall number
Z-0867-2020
Recall status
Terminated
CDRH record ID
178258
FDA classified
2020-01-27
Recall event number
84411

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗