Skip to content

Public FDA record

Z-0868-03: Stryker Secure II hospital bed; model 3002.

Recall record Z-0868-03 for Stryker Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0868-03
Company in FDA record
Stryker Medical
FDA classification date
Jun 12, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A wire may break at the crimp and present a potential electric shock hazard.

Source facts

Action
The firm sent service representatives to the receiving hospitals, beginning on 5/1/02, and the beds were corrected in place.
Initiated
2003-05-01
Root cause
Other
Terminated
2003-06-11
Distribution
Pennsylvania and Kansas
Recall number
Z-0868-03
Recall status
Terminated
CDRH record ID
27511
FDA classified
2003-06-12
Recall event number
26325

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗