This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
Source facts
Action
The recalling firm issued Dear Doctor letters dated 11/16/2018 explaining the reason for the voluntary correction and pointing out a correction was previously initiated in June, 2017, related to this issue for devices distributed prior to October, 2015.
Initiated
2018-11-16
Root cause
Manufacturing material removal
Terminated
2022-12-12
Distribution
Distribution US nationwide, including Puerto Rico. There was no foreign/military/government consignees.
Recall number
Z-0868-2020
Recall status
Terminated
CDRH record ID
177027
FDA record created
2020-01-27
Recall event number
84029