Open original FDA record ↗Recall record
- FDA identifier
- Z-0868-2021
- Company in FDA record
- Laerdal Medical (Suzhou) Co., Ltd.
- FDA record creation date
- Jan 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Component failures may cause the unit to lose all suction. Users will need to obtain alternate suction devices or methods, leading to a delay in treatment and, in worst case scenario, potential suffocation if alternative suction or ventilation method is not administered by the user.
Source facts
- Action
- On December 18, 2020, the firm sent Urgent Medical Device Recall letters to affected customers. Customers were informed that the service life of the LCSU 4 may be reduced. Customers were asked to identify and secure affected units. Customers were told that they could continue to use their LCSU 4 unit until they receive a replacement unit, but were provided with safety measures: - Follow local protocols for back-up suction devices. Do not repeatedly block the patient tube for more than 10 seconds at a time at vacuum levels greater than 300 mmHg. - Only use canisters, tubes and accessories that are approved for the device by Laerdal Medical. - When adjusting the vacuum level: When the desired vacuum level is reached; unblock the patient tube immediately. - During patient treatment do not occlude or block the patient tube by intentional bending or squeezing. The patient tube should only be temporarily occluded by suctioning matter. - When performing the Device Test: When the vacuum reaches minimum 500 mmHg immediately release the blocking. - If your affected LCSU 4 experiences a loss of peak suction, please report this to your local Laerdal Sales office, for US: 877-LAERDAL (523-7325), customerservice@laerdal.com or, for Canada: (416) 298-9600, savelives@laerdal.com Customers must complete the included reply form to acknowledge receipt of the recall notice and to identify any units needing replacement, and return the information to Stericycle.
- Initiated
- 2020-12-18
- Root cause
- Device Design
- Terminated
- 2023-12-11
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Recall number
- Z-0868-2021
- Recall status
- Terminated
- CDRH record ID
- 184340
- FDA record created
- 2021-01-15
- Recall event number
- 86702