Firmware defect. Under certain conditions a strip rack is no longer in the initial position after a motor overload of the barcode scanner PosID2.
Source facts
Action
Tecan US notified consignees by registered letter on/about 3/13/2003. International accounts were notified via e-mail 3/13/2003 by Tecan Switzerland. Consignees were advised to contact the firm''s Technical Support line to coordinate the PosID firmware upgrade.
Initiated
2003-02-27
Root cause
Other
Terminated
2006-04-06
Distribution
Instruments requiring upgrade were distributed to various clinical laboratories and blood banks Nationwide and World wide .
Recall number
Z-0869-03
Recall status
Terminated
CDRH record ID
26603
FDA classified
2003-05-29
Recall event number
25850