Z-0869-2012: STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Clea

Recall record Z-0869-2012 for Steris Corporation Hopkins Facility. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0869-2012
Company in FDA record
Steris Corporation Hopkins Facility
FDA classification date
Jan 25, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature. Failure to meet these criteriae may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant Processing System. The firm intends to perform a software field correction which

Source facts

Action
The firm, STERIS Corp. prepared an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated December 8, 2011. The firm intends to send the letter to its customers upon STERIS' receipt of 510(k) clearance from FDA. The letter includes a description of the Product-SYSTEM Liquid Chemical Sterilant Processing System, problem and actions taken. The customers were instructed to in case of an incomplete or cancelled cycle, reprocess devices in the cycle following the directions in the Operator manual. The letter also states that STERIS Corporation will install updated software on the affected system, and will be providing the customers with a new Operator Manual reflecting the revised wording resulting from the installation of the new software. For further information or if they have any questions regarding the firms visit to their facility, contact STERIS Field Service Dispatch at 1-800-333-8828. If you have questions regarding this matter, contact Director Low Temperature Sterilization, at 440-392-7455.
Initiated
2011-12-07
Root cause
Software design
Terminated
2014-05-13
Distribution
Worldwide distribution: USA (nationwide) and countries of: Bahrain, Hong Kong, Japan and South Korea.
Recall number
Z-0869-2012
Recall status
Terminated
CDRH record ID
106373
FDA classified
2012-01-25
Recall event number
60742

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗