Open original FDA record ↗Recall record
- FDA identifier
- Z-0870-2007
- Company in FDA record
- General Electric Med Systems LLC
- FDA classification date
- Jun 5, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control room) or in-room InnovaCentral/Touch-Screen
Source facts
- Action
- A letter dated Jan. 18, 2007, was sent to consignees describing the problem, product ,safety instructions and contact information. Software upgrades were promised for correction.
- Initiated
- 2007-03-26
- Root cause
- Other
- Terminated
- 2008-10-21
- Distribution
- Worldwide Distribution: USA states including-AZ,CA,CO,DE,FL,GA,IL,IN, MD,MN,MO,NJ,NV,NY,OH,OK,OR,PA,SC,SD,TX, UT, VA,WA,WI, and countries of Canada, Belgium, China, Czech Republic, Finland, France, Germany, Hungary, India, Isreal, Italy, Japan, Korea, New Zealand, Norway, Russia, Spain, and United Kingdom.
- Recall number
- Z-0870-2007
- Recall status
- Terminated
- CDRH record ID
- 51382
- FDA classified
- 2007-06-05
- Recall event number
- 37723