Open original FDA record ↗Recall record
- FDA identifier
- Z-0870-2010
- Company in FDA record
- Remel, Inc
- FDA classification date
- Feb 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is an error in Version 1.0.77 of the electronic code compendium (ERIC) software. The error is limited to the database corresponding to the R8311007 RapID Yeast Plus System. The database erroneously includes Candida dubliniensis as well as displays an incorrect temperature range as an accessory test for codes with an overlap between Candida albicans and Candida dubliniensis. When using the
Source facts
- Action
- The firm issued an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" dated October 29, 2009. After completion of the firm's investigation and risk assessment, Remel decided to notify customers of this situation and quarantine any remaining units on hand. The notification provided the consignees with instructions for obtaining and downloading the software patch. A response form was included with the recall letter and the consignee was asked to complete the form and return it to Remel. The consignees should contact the Technical Services Department at 800-255-6730 (USA) or 913-888-0939 (International), if they have inquiries concerning this advisory.
- Initiated
- 2009-10-28
- Root cause
- Software design
- Terminated
- 2010-03-03
- Distribution
- Worldwide Distribution: USA, Europe and Canada
- Recall number
- Z-0870-2010
- Recall status
- Terminated
- CDRH record ID
- 86536
- FDA classified
- 2010-02-26
- Recall event number
- 53800