Z-0870-2020: EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emit

Recall record Z-0870-2020 for Energetiq Technology Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0870-2020
Company in FDA record
Energetiq Technology Inc
FDA record creation date
Jan 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Source facts

Action
On October 28, 2019 Recall notification letters were sent to all customers of record that purchased the affected unit. The letter identified the affected product, problem and actions to be taken. Pending resolution of this issue we would advise all users if EQ-4001 units to follow the Operational Requirements that are attached to this letter. Specifically DO NOT REMOVE THE LASER COLLIMATOR FROM THE LAMP WHILE THE SYSTEM IS OPERATING. We have also enclosed a safety label and bolt caps with installation instructions. The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of H Health and Human Services, the details of which will be included in a subsequent communication to you.
Initiated
2019-09-27
Root cause
Device Design
Distribution
Worldwide Distribution.
Recall number
Z-0870-2020
Recall status
Open, Classified
CDRH record ID
178195
FDA record created
2020-01-29
Recall event number
84378

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗