Open original FDA record ↗Recall record
- FDA identifier
- Z-0870-2021
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Jan 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Boston Scientific (BSC) is retrieving five Model 7060 ACUITY cutters distributed in the US that did not undergo the applicable sterile processing procedure.
Source facts
- Action
- The firm initiated the recall by hand delivered letter. The letter explained the issue and requested the following actions be taken: "1. Immediately segregate all affected product to ensure that it will not be used. 2. Your local Boston Scientific sales representative will retrieve affected product. 3. Complete the enclosed Reply Verification Tracking Form. 4. If affected product has been used, please inform the implanting physician for appropriate patient management." "If you have further questions, please contact your local Boston Scientific representative or United States Technical Services at 1-800-227-3422."
- Initiated
- 2020-12-15
- Root cause
- Process control
- Terminated
- 2023-08-14
- Distribution
- US Nationwide distribution including in the states of Tulsa, OK.
- Recall number
- Z-0870-2021
- Recall status
- Terminated
- CDRH record ID
- 185073
- FDA record created
- 2021-01-15
- Recall event number
- 87033