Open original FDA record ↗Recall record
- FDA identifier
- Z-0871-2015
- Company in FDA record
- Baxter Corporation Englewood
- FDA classification date
- Dec 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Baxter Corporation initiated a recall of two lots of Self-Righting Luer Lock Tip Caps because they were distributed prior to closure of the validation.
Source facts
- Action
- Baxter sent an Urgent Product Recall letter via USPS to all affected customers dated December 8, 2014.The letter identified the product the problem and the action needed to be taken by the customer. Customers are instructed to locate and remove all affected product from their facility, and to contact Baxter Healthcare to arrange for return and credit.Returned products will be destroyed./discarded. If you have questions regarding the content of the communication, please call The Center for One Baxter at (1-800) 422-9837, between the hours of 8:00am and 5:00pm Central time.
- Initiated
- 2014-12-04
- Root cause
- Process control
- Terminated
- 2016-11-30
- Distribution
- Worldwide Distribution - US Distribution to the states of UT, TX, and IL, and to the countries Austria, Germany, Spain, Switzerland, and Netherlands. No VA/military/govt consignees.
- Recall number
- Z-0871-2015
- Recall status
- Terminated
- CDRH record ID
- 131999
- FDA classified
- 2014-12-24
- Recall event number
- 69934