Open original FDA record ↗Recall record
- FDA identifier
- Z-0871-2026
- Company in FDA record
- Paragon 28, Inc.
- FDA classification date
- Dec 2, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
Source facts
- Action
- On 10/13/2025, recall notices were emailed to distributors who were asked to do the following: 1) Return affected devices to the firm using return label. 2) Ensure that in instances where you have distributed this affected product, that you forward this Recall Notification to your customers. 3) Replacements are available by contacting Orders@paragon28.com Complaints can be directed to complaints@paragon28.com
- Initiated
- 2025-10-13
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
- Recall number
- Z-0871-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216972
- FDA classified
- 2025-12-02
- Recall event number
- 97891