Skip to content

Public FDA record

Z-0872-2021: REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product U

Recall record Z-0872-2021 for Medtronic Neurosurgery. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0872-2021
Company in FDA record
Medtronic Neurosurgery
FDA record creation date
Jan 15, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction

Source facts

Action
On 12/16/2020, Medtronic sent a "URGENT: MEDICAL DEVICE RECALL" notification to affected customers via FedEx. In addition, to notifying consignees about the recall, the firm ask consignees to take the following actions: 1. Identify, segregate, and quarantine affected valves within your inventory. 2. Contact Medtronic to return affected valves and receive replacement(s). See instructions below. 3. Please complete and return the customer confirmation form, even if you do not have any affected valves. 4. Return the form via email to RS.NavFCA@Medtronic.com or via fax to 651-367-7075 within 30 days of receipt.
Initiated
2020-12-16
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IL, KS, KY, LA, MA, MI, MO, NV, NM, NY, ND, OH, OK, TN, VA and the countries of Albania, Australia, Brazil, Cyprus, Ireland, Mexico, Qatar, Romania, Russian Federation, Serbia, South Africa, Spain, United Kingdom.
Recall number
Z-0872-2021
Recall status
Open, Classified
CDRH record ID
185159
FDA record created
2021-01-15
Recall event number
87069

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗