Open original FDA record ↗Recall record
- FDA identifier
- Z-0872-2022
- Company in FDA record
- B Braun Medical Inc
- FDA record creation date
- Apr 15, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
Source facts
- Action
- On February 17, 2022, the firm, B.Braun Medical Inc., mailed out Recall Customer Notification Letters to affected consignees. Distributors should notify affected customers. Customers should identify any affected products within the spare parts inventory of their facility, cease use, and quarantine the product. Do not destroy any affected product. If any machine exhibited repeated Alarm Code 1026 - "UF balance? Air leakage in dialyzer coupling" alarms, customer should discontinue therapy on the device and immediately quarantine the machine until it can be inspected and any impacted sensors replaced. Customers should inspect ALL Dialog+ and Dialog+ Evolution dialysis machines in their facility which may have been serviced with potentially impacted sensors. Once the firm receives the customer acknowledgement form, a representative will contact the customer on how to return impacted sensors and/or to schedule a time to inspect equipment. Customers with questions should contact B. Braun Avitum Technical Services at 1-800-21-0445.
- Initiated
- 2022-01-28
- Root cause
- Process control
- Terminated
- 2024-05-15
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0872-2022
- Recall status
- Terminated
- CDRH record ID
- 192434
- FDA record created
- 2022-04-15
- Recall event number
- 89784