Open original FDA record ↗Recall record
- FDA identifier
- Z-0874-2015
- Company in FDA record
- Ebi, Llc
- FDA classification date
- Dec 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.
Source facts
- Action
- The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated 11/11/2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately locate and remove from circulation the items consigned/loaned to your account as identified; carefully follow the instructions on the enclosed "Response Form"; complete and return Response Form via fax to 303.501.8444, and if you have further distributed these items to medical facilities, you MUST notify them of the actions. Questions related to this notice should be directed to the Sr. Quality Engineer at 518-859-3487 Monday through Friday, 8 am to 4 pm ET.
- Initiated
- 2014-11-11
- Root cause
- Device Design
- Terminated
- 2015-04-15
- Distribution
- US Nationwide Distribution.
- Recall number
- Z-0874-2015
- Recall status
- Terminated
- CDRH record ID
- 132064
- FDA classified
- 2014-12-24
- Recall event number
- 69981