Skip to content

Public FDA record

Z-0874-2015: Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sl

Recall record Z-0874-2015 for Ebi, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0874-2015
Company in FDA record
Ebi, Llc
FDA classification date
Dec 24, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.

Source facts

Action
The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated 11/11/2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately locate and remove from circulation the items consigned/loaned to your account as identified; carefully follow the instructions on the enclosed "Response Form"; complete and return Response Form via fax to 303.501.8444, and if you have further distributed these items to medical facilities, you MUST notify them of the actions. Questions related to this notice should be directed to the Sr. Quality Engineer at 518-859-3487 Monday through Friday, 8 am to 4 pm ET.
Initiated
2014-11-11
Root cause
Device Design
Terminated
2015-04-15
Distribution
US Nationwide Distribution.
Recall number
Z-0874-2015
Recall status
Terminated
CDRH record ID
132064
FDA classified
2014-12-24
Recall event number
69981

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗