Z-0874-2022: Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion

Recall record Z-0874-2022 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0874-2022
Company in FDA record
Philips North America Llc
FDA record creation date
Apr 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

Source facts

Action
Urgent Medical Device Correction notification letters dated 2/2/22 were distributed to customers. 1. What is the problem and under what circumstances it can occur Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient. When the problem occurs, the area of the boosted images remains black, and the small X-ray viewer in the upper right corner will show a single static image of the previous run. The correct patient information is shown on the Philips Allura/ Azurion system. This defect is intermittent and is caused by a race condition when the software does not properly handle all the "process requests". This problem has been identified through the investigation of 2 (two) customer complaints. 2. What is the hazard/harm assoclated with this issue The incorrect image displayed to the user could lead to incorrect treatment. If the problem occurs, the Stent8oost Live application will have to be restarted (see action 4), causing a delay in the procedure. To date, Philips has not received any reports of harm associated with this problem. 3. Affected products and how to identify them The Philips StentBoost Live R2.0 used with Philips Allura and Azurion systems is affected by this issue. The software version of the Philips StentBoost Live application can be seen by clicking on the "About" box displayed when the application is opened (see Fig. 1 and 2).
Initiated
2022-02-02
Root cause
Under Investigation by firm
Distribution
Global Distribution. US Nationwide.
Recall number
Z-0874-2022
Recall status
Open, Classified
CDRH record ID
192431
FDA record created
2022-04-06
Recall event number
89787

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗