Skip to content

Public FDA record

Z-0875-2015: Trilogy Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either

Recall record Z-0875-2015 for Zimmer Manufacturing B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0875-2015
Company in FDA record
Zimmer Manufacturing B.V.
FDA classification date
Dec 24, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which

Source facts

Action
The firm, Zimmer, sent an "URGENT MEDICAL DEVICE RECALL-LOT SPECIFIC" memo dated 12/4/2014 to its customers. The memo described the product, problem and actions to be taken. The customers were instructed to locate all product from the listed lot and quarantine them immediately; carry out a physical count of all affected product in inventory; complete and return the Inventory Return Certification Form via email to: corporatequality.postmarket@zimmer.com on or before 12/12/2014; return the recalled product along with Inventory Return Certification Form on or before 12/12/2014; keep and maintain a copy for your records; and notify Zimmer of any hospitals that you have further distributed the affected product to and complete spreadsheet template provided. After reviewing this notification you have further questions or concerns, call the customer call center at 1-800-348-2759 between 8:00 am and 5:00 pm EST.
Initiated
2014-12-04
Root cause
Under Investigation by firm
Terminated
2015-06-18
Distribution
Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.
Recall number
Z-0875-2015
Recall status
Terminated
CDRH record ID
131584
FDA classified
2014-12-24
Recall event number
69814

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗