Open original FDA record ↗Recall record
- FDA identifier
- Z-0876-06
- Company in FDA record
- Dade Behring , Inc.
- FDA classification date
- May 19, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
Source facts
- Action
- Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.
- Initiated
- 2006-04-03
- Root cause
- Other
- Terminated
- 2006-08-07
- Distribution
- Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.
- Recall number
- Z-0876-06
- Recall status
- Terminated
- CDRH record ID
- 45345
- FDA classified
- 2006-05-19
- Recall event number
- 35112