Z-0876-2012: Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply co

Recall record Z-0876-2012 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0876-2012
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jan 26, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August 15, 2011 has a higher standby current resulting in the battery depleting prematurely.

Source facts

Action
Carefusion sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated September 13, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A Customer Reply Card was enclosed for customers to complete and return to the firm. Customers were instructed to call Carefusion Support Center at (888) 562-6018, 7 am-5 pm (Pacific) for any recall related questions.
Initiated
2011-09-13
Root cause
Nonconforming Material/Component
Terminated
2012-01-31
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CT, FL, IA, IL, MI, MO, NJ, NY, OH, SC, SD, and TX. and the country of Australia.
Recall number
Z-0876-2012
Recall status
Terminated
CDRH record ID
106751
FDA classified
2012-01-26
Recall event number
60904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗