Z-0876-2021: Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and i

Recall record Z-0876-2021 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0876-2021
Company in FDA record
Steris Corporation
FDA record creation date
Jan 15, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.

Source facts

Action
On December 11, 2020, the firm distributed Urgent Medical Device Correction letters to customers. The letter informed customers that heating elements in the drying chamber may overheat, eventually resulting in smoke and/or fire. The firm has developed an automatic shut off of the power to prevent the heating elements from overheating if this malfunction were to occur. The unit would remain "ON" in an alarm state, informing the user that service is required. All affected customers will be contacted by a STERIS service technician to arrange an onsite inspection of their unit(s) and perform the correction. Users can continue to use their devices prior to completion of the correction. Users are reminded to follow the instructions outlined in the Operator Manual should the unit alarm "Drying Temperature Too High" by aborting the cycle, turning off power to the unit, and contacting STERIS. If you have questions regarding this matter, please contact Sarah Brown, Senior Product Manager, at 440-392-7571, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
Initiated
2020-12-11
Root cause
Component design/selection
Terminated
2022-06-21
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Bulgaria, Canada, Chile, China, Ecuador, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of Malaysia, Mexico, New Zealand, Panama, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela.
Recall number
Z-0876-2021
Recall status
Terminated
CDRH record ID
184858
FDA record created
2021-01-15
Recall event number
86937

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗