Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.
Source facts
Action
Care Innovations sent a "DEVICE CORRECTION NOTIFICATION" letter dated 12/20/12 to all affected customers. The letter identified the product and problem. Correction was performed remotely by the firm. Contact the firm at 917-232-4775 for questions regarding this notice.
Initiated
2012-12-16
Root cause
Software design
Terminated
2013-02-26
Distribution
Nationwide Distribution.
Recall number
Z-0877-2013
Recall status
Terminated
CDRH record ID
115109
FDA classified
2013-02-23
Recall event number
63957