Skip to content

Public FDA record

Z-0877-2015: Synthes Inflation System (03.804.4135) Product Usage: Inflation system is used during kyphoplasty or ot

Recall record Z-0877-2015 for PEROUSE MEDICAL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0877-2015
Company in FDA record
PEROUSE MEDICAL
FDA classification date
Dec 29, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Additional instructions for use due to inability to raise the pressure above 10 atm during inflation phase may result in an incomplete inflation or delay in procedure

Source facts

Action
Perouse Medical notified Distributors by Safety Alert Letter sent on October 06, 2014 and and October 07, 2014 by e-mail. The letter identified the defective products and lot numbers and provided a description of the malfunction, risk and mitigation statements. US distributors have been requested to provide the required information to their customers. In addition, the firm has updated device instructions for use. The firm is requiring its customer to fill out a safety notice acknowledgement of receipt form.
Initiated
2014-10-06
Root cause
Device Design
Terminated
2017-03-24
Distribution
US Nationwide Distribution in the states of AZ and CO.
Recall number
Z-0877-2015
Recall status
Terminated
CDRH record ID
131893
FDA classified
2014-12-29
Recall event number
69879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗