Z-0877-2021: uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001

Recall record Z-0877-2021 for Roche Molecular Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0877-2021
Company in FDA record
Roche Molecular Systems, Inc.
FDA record creation date
Jan 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.

Source facts

Action
On November 23, 2020, the firm initiated its recall. The firm distributed Urgent Medical Device Correction letters to affected customers using uPath software version 1.1 IVD. Until uPath Enterprise software version 1.1.1 patch is installed, customers using uPath Enterprise software version 1.1 should not analyze slides with different scan magnifications in Split View. In order to reduce the risk of this issue impacting patient care, customers should follow their local procedures and policies regarding retrospective re-review of cases. In particular, it is recommended that cases involving tumor measurements (e.g., melanoma, etc.) be considered for re-review, especially if the original diagnosis included a reported measurement (e.g., Breslow depth in melanoma, etc.). Pathologists need to assess the risk of this measurement error in the context of local pathology reporting processes and clinical decision making. Re-review should be limited to the following conditions, all of which are required for the issue to occur: 1. Slides were viewed in uPath Enterprise software version 1.1, 2. The tumor measurement was done in the Split View viewing mode, and 3. The slides within the Split View viewing mode were scanned at two different magnifications. Only under these conditions is a re-review and repeat tumor measurement recommended.
Initiated
2020-11-23
Root cause
Software design
Terminated
2023-05-22
Distribution
Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.
Recall number
Z-0877-2021
Recall status
Terminated
CDRH record ID
185012
FDA record created
2021-01-17
Recall event number
86996

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗