Open original FDA record ↗Recall record
- FDA identifier
- Z-0877-2023
- Company in FDA record
- Covidien LP
- FDA classification date
- Jan 12, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Source facts
- Action
- On December 7, 2022, Medtronic began notifying customers who had received the affected products. Customers were instructed to quarantine, discontinue use, and return unused products. Customers should forward the notification to those who need to be aware within the organization, and to any location where the devices have been distributed.
- Initiated
- 2022-12-07
- Root cause
- Process design
- Distribution
- Worldwide distribution. US nationwide including Guam, USVI, and Puerto Rico; Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, Italy, Jamaica, Kenya, Malaysia, Malta, Mauritius, Mexico, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Uganda, United Arab Emirates, and United Kingdom.
- Recall number
- Z-0877-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197255
- FDA classified
- 2023-01-12
- Recall event number
- 91206