Skip to content

Public FDA record

Z-0878-2015: ev3 Protege Rx Tapered Carotid Stent System, SECX-10-7-40-135, Sterile EO, Rx only. Indicated for the t

Recall record Z-0878-2015 for Ev3, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0878-2015
Company in FDA record
Ev3, Inc.
FDA classification date
Dec 29, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two production lots of the Protege Rx Tapered Carotid Stent Systems (SECX-10-7-40-135, Lot # 9922452 & SEPX-8-6-40-135, Lot # 9922795) because of a product labeling error.

Source facts

Action
Consignees were sent on 12/4/2014 a Covidien "Urgent Field Safety Notice" letter dated 4th December 2014. The letter described the problem and the product involved in the recall. Advised consignees to quarantine, discontinue use of the device and return the product to Covidien. They also requested consignees to complete the "Recall Verification Form" and return it to Covidien. For questions, customers can contact their local Covidien Representative.
Initiated
2014-12-04
Root cause
Packaging process control
Terminated
2015-07-17
Distribution
International Only Distribution -- RUSSIA, ROMANIA, UNITED KINGDOM, GERMANY, ITALY, SPAIN, FRANCE, TURKEY, POLAND, and JAPAN.
Recall number
Z-0878-2015
Recall status
Terminated
CDRH record ID
132042
FDA classified
2014-12-29
Recall event number
69965

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗