Open original FDA record ↗Recall record
- FDA identifier
- Z-0878-2021
- Company in FDA record
- Fujifilm Medical Systems U.S.A., Inc.
- FDA record creation date
- Jan 17, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for the wrong patient information may be displayed in the viewer or PowerJacket. Additionally, the incorrect max standard uptake values (SUV) for PET ad CT studies may be returned for Siemens Modality.
Source facts
- Action
- Letter was sent to all affected customers on 09/11/2020 stating the following: There have been no reported patient injuries associated with these issues. In an effort to ensure the highest level of customer satisfaction, FUJIFILM will offer an upgrade to Synapse PACS 5.7.200US to resolve these issues. Until your upgrade is scheduled and completed, you can continue to safely use Synapse PACS by following these additional instructions: 1. When selecting a patient off the PowerJacket or work list, please ensure that the patient information displayed in the viewer matches the desired patient and matches the PowerJacket. 2. If using a Siemens Modality, Max SUVs should not be used when rendering a clinical decision until the issue is corrected. Please complete and return the Field Action Verification Form attached on page 3 of this communication. Providing the response with the information requested is essential for ensuring appropriate action is taken.
- Initiated
- 2020-09-11
- Root cause
- Device Design
- Distribution
- Software was distributed to medical facilities nationwide throughout the U.S.
- Recall number
- Z-0878-2021
- Recall status
- Open, Classified
- CDRH record ID
- 183942
- FDA record created
- 2021-01-17
- Recall event number
- 86532