Skip to content

Public FDA record

Z-0878-2023: VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact

Recall record Z-0878-2023 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0878-2023
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jan 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results

Source facts

Action
Update 12/15/22: Ortho Clinical Issued Further Corrective Action: Resolution To Potential For Negative Bias When Using Vitros Immunodiagnostic Products Intact Pth Reagent Pack. Letter states: Beginning with Lot 1676 (and subsequent lots), VITROS iPTH Reagent Pack has undergone a performance change in the math associated with the internal master calibration to remove the negative bias observed in previous affected lots. This change brings the performance of VITROS iPTH Reagent Pack back to expected levels. The next lot of VITROS Immunodiagnostic Products iPTH Controls to support this change will be Lot 1090. Please note that Lots 1676 and above of iPTH Reagent Pack will only be supported by Lots 1090 and above of VITROS iPTH Controls. Beginning with Lot 1676 (and subsequent lots) of VITROS iPTH Reagent Pack, customers will no longer be required to adjust their laboratory s reference interval when performing patient sample testing. When using VITROS iPTH Reagent Pack intraoperatively, any unaffected lot may once again be used for patient sample testing. To maintain the expected performance of this product, after switching to Lot 1676 (and subsequent lots) customers are advised not to switch back to any of the previous affected lots of VITROS iPTH Reagent Pack. Ortho has also provided new control values for the non-Ortho controls listed beginning on page 2. " Please note that these new control values are available for use with Lots 1676 and above only. " The non-Ortho control values provided in the previous communication (Ref. CL2022-275) are acceptable for use with VITROS iPTH Reagent Pack Lots 1610, 1621, 1630, 1640, 1645, 1650, and 1670. " The original non-Ortho control values, assigned during release, are acceptable for use with Lot 1660 and Lots prior to 1610. " New control lots will be suitable for use with Lot 1676 and above only. The new lots of VITROS iPTH Reagent Pack and VITROS iPTH Controls are available now. To ensure product availability, Ortho has impl
Initiated
2022-11-17
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Bermuda HM12, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Recall number
Z-0878-2023
Recall status
Open, Classified
CDRH record ID
197254
FDA classified
2023-01-04
Recall event number
91205

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗