A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse, the pouch package of 5 of the 10 individually packaged cannula had not been sealed along one edge. Thus the pouch was open and the device could fall out of the pouch and on to the floor when retrieved from the shelf box. The cannulas were not used on any patients, as the unsealed packages
Source facts
Action
A recall was initiated on 10/04/2012 and a letter was sent out to the distributor with a Fax-back Acknowledgment Form. The letter described the issue and provided recommendations and a contact number.
Initiated
2012-10-04
Root cause
Packaging
Terminated
2013-05-23
Distribution
International distribution to the country of Australia.
Recall number
Z-0879-2013
Recall status
Terminated
CDRH record ID
115670
FDA classified
2013-02-26
Recall event number
64250