Skip to content

Public FDA record

Z-0880-2021: Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized c

Recall record Z-0880-2021 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0880-2021
Company in FDA record
Abbott Point Of Care Inc.
FDA record creation date
Jan 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Source facts

Action
Abbott Point of Care notified Customers via email beginning on 22 December 2020. Customer letters will begin shipping via FedEx on 28-December-2020 for delivery on or around 29-December-2020. Letter states reason for recall, health risk and action to take: While Abbott Point of Care continues to investigate this issue, if reporting ionized calcium is required, it is recommended that the i-STAT CG8+ and EG7+ cartridges be used only: within 3 days of removal from refrigerated storage for lot numbers between *20100 and *20339 within 7 days of removal from refrigerated storage for lot numbers *20340 and above Where *= lot letter A, K, L, M, N, W and Y If your facility is unable to manage to the recommended action outlined above, Abbott Point of Care recommends that your facility disable and not report the ionized calcium test on the i-STAT CG8+ and EG7+ cartridges and use an alternate i-STAT cartridge or method for reporting ionized calcium. Please see additional information below on how to disable the ionized calcium test for these cartridges. Please confirm receipt and understanding of this communication by responding to the business reply form included with this letter. Additional information regarding potential solutions will be provided separately by APOC. If you have forwardedany i-STAT CG8+ and EG7+ cartridges to another facility, we request that you please provide a copy of this letter to them. Your Abbott Point of Care sales and support team will be contacting you to review interim solutions for the challenges this communication may bring. If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 1-844-256-9531, or via email at apoc_productinformation@abbott.com or visit the Abbott Point of Care website www.pointofcare.abbott. Additionally, please report any questions or concerns about the performance of the ionized calcium test on the i-STAT CG8+ and EG7+ cartridges to Abbott Point of Care Tec
Initiated
2020-12-22
Root cause
Under Investigation by firm
Terminated
2023-09-01
Distribution
Nationwide
Recall number
Z-0880-2021
Recall status
Terminated
CDRH record ID
185166
FDA record created
2021-01-19
Recall event number
87073

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗