Recall record Z-0880-2023 for Altimate Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0880-2023
Company in FDA record
Altimate Medical, Inc.
FDA classification date
Jan 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device manufactured with a component with affects the foldability of the frame, not allowing it to be secured by the frame strap in the folded position.

Source facts

Action
Altimate Medical notified its sole consignee by telephone on 11/30/2022 and followed with an Urgent Medical Device Recall letter on 12/06/2022. The notice stated the problem and requested the following action: "- Acknowledge receipt of this notification. Complete and return the attached Acknowledgement form. - Inform the customer not to use the chair if this has not already done. - Provide replacement chair 90071504001 under RMA000737 to AMI - Inform Altimate when the replacement chair is provided to the customer. Complete and return the attached Verification form."
Initiated
2022-11-30
Root cause
Component change control
Terminated
2023-08-31
Distribution
US Nationwide distribution in the state of Texas.
Recall number
Z-0880-2023
Recall status
Terminated
CDRH record ID
197504
FDA classified
2023-01-04
Recall event number
91335

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗