The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
Source facts
Action
On 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
Initiated
2020-12-23
Root cause
Device Design
Terminated
2023-04-13
Distribution
Nationwide distribution.
Recall number
Z-0882-2021
Recall status
Terminated
CDRH record ID
185324
FDA classified
2021-01-21
Recall event number
87095