Open original FDA record ↗Recall record
- FDA identifier
- Z-0883-04
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Jul 20, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The tension pin was not installed in spreader bar pivot bolt to prevent the pivot bolt from migrating out of the spreader bar, allowing the spreader bar to detach from the lift.
Source facts
- Action
- All affected end users were given a copy of the Customer Advisory Notice dated 1/30/03, during the Arjo salesmen visits to inspect the Minstrel lifts for the presence of the tension pin in the pivot bolt of the lift. The advisory informed the user of the incident where the spreader bar detached from the scale unit. The end user was requested to read the advisory and sign the bottom portion of it certifying that they read the advisory and that the Arjo salesman inspected their units for the presence of the tension pin, retaining a copy of the signed advisory notice and the inspeciton record.
- Initiated
- 2003-01-30
- Root cause
- Other
- Terminated
- 2004-05-12
- Distribution
- Nationwide. District of Columbia, New York, Washington, Iowa, Florida, North Carolina, Louisiana, Virginia, South Carolina, Minnesota and Colorado
- Recall number
- Z-0883-04
- Recall status
- Terminated
- CDRH record ID
- 32647
- FDA classified
- 2004-07-20
- Recall event number
- 28866