Reason stated in the source

The tension pin was not installed in spreader bar pivot bolt to prevent the pivot bolt from migrating out of the spreader bar, allowing the spreader bar to detach from the lift.

Source facts

Action
All affected end users were given a copy of the Customer Advisory Notice dated 1/30/03, during the Arjo salesmen visits to inspect the Minstrel lifts for the presence of the tension pin in the pivot bolt of the lift. The advisory informed the user of the incident where the spreader bar detached from the scale unit. The end user was requested to read the advisory and sign the bottom portion of it certifying that they read the advisory and that the Arjo salesman inspected their units for the presence of the tension pin, retaining a copy of the signed advisory notice and the inspeciton record.
Initiated
2003-01-30
Root cause
Other
Terminated
2004-05-12
Distribution
Nationwide. District of Columbia, New York, Washington, Iowa, Florida, North Carolina, Louisiana, Virginia, South Carolina, Minnesota and Colorado
Recall number
Z-0883-04
Recall status
Terminated
CDRH record ID
32647
FDA classified
2004-07-20
Recall event number
28866

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗