Skip to content

Public FDA record

Z-0883-2021: A calculator/data processing module for clinical use is an electronic device intended to store, retrieve,

Recall record Z-0883-2021 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0883-2021
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Jan 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software error results in the unintentional removal of the serum-indices flag that would otherwise prevent the release of results. The Cobas Inifinity has an auto-verification feature to hold results for manual review when they meet specific criteria. The serum indices flag is used for serum indices tests that are performed to assess the quality of the sample (e.g. hemolysis, icterus, and lipemia). Normally, when the Cobas Infinity receives a result for a test that is serum indices-sensitive, it flags the result and the software then waits for the results of the serum indices tests before validating or rejecting the test result. Roche has discovered a software error that under specific conditions causes the flag to be incorrectly removed. This allows for the possibility that a sample of poor quality may return an unreliable/incorrect test result that is mistakenly reported to the health care provider and/or patient without the proper disclaimer that the result is based on a sample of poor quality.

Source facts

Action
Roche issued Urgent Medical Device Correction, (UMDC TP-01176) via UPS Ground (receipt required) on 12/17/20. The UMDC TP-01176 will also post to the diagnostics.roche.com website. Letter states reason for recall, health risk and action to take: Complete the enclosed faxback form, TP-01177, and return it following the instructions on the form even if you are not using the Cross SI option. " If you are using the Cross SI option, refer to the Workarounds section of this Urgent Medical Device Correction (UMDC) until the Service Patch for your cobas infinity central lab is available for this issue. " File this UMDC for future reference Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC
Initiated
2020-12-17
Root cause
Software design
Terminated
2023-05-19
Distribution
US Nationwide
Recall number
Z-0883-2021
Recall status
Terminated
CDRH record ID
185110
FDA record created
2021-01-22
Recall event number
87040

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗