Z-0883-2026: Lumify Diagnostic Ultrasound System models : 795005, 795161, 989605449841,

Recall record Z-0883-2026 for Philips Ultrasound, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0883-2026
Company in FDA record
Philips Ultrasound, LLC
FDA classification date
Dec 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining.

Source facts

Action
On November 1, 2025, Philips issues an Urgent Medical Device Recall (Correction) Notification to affected consignees. Phillips ask consignees to take the following actions: 1. Utilize an Android based device with your Lumify system. 2. Lightning port devices are not affected and may continue to be used as normal. 3. If you use an 05 based device, do not update your OS device to software version lax. 4. Turn off 05 auto-updates on USB host devices for Lumify by following the below steps: 5. Place this Notice with the documentation of the affected Ultrasound system(s); ensure the notice is in a place likely to be viewed by device operators. 6. Circulate this Notice to all users of the affected device(s) so that they are aware of the issue. 7. Complete the Acknowledgment form indicating you have read and understand the recall.
Initiated
2025-10-31
Root cause
Software design
Distribution
Worldwide US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR and the countries of Argentina, Armenia, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Dominican Rep, Ecuador, El Salvador, Estonia, France, Germany, Greece, Guatemala, Honduras , Hungary, Iceland, India, Indonesia, Israel, Italy, Jordan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Singapore, Slovakia, South Africa, South Korea, Spain, St. Pier/Miquel., Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Tunisia, T¿rkiye, United Kingdom, Uruguay, Vietnam.
Recall number
Z-0883-2026
Recall status
Open, Classified
CDRH record ID
216707
FDA classified
2025-12-04
Recall event number
97843

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗