Units are defective in that the identification and certification labels were inadequate or missing. The beam limiting device was not appropriate for general purpose radiographic procedures and the beam limiting device could ot collimate down to 125 square centimeters or less as required.
Source facts
Action
Field Service Engineers will install a dual fixed-size colimator and console label and a new user manual will be provided.
Initiated
2004-06-11
Root cause
Other
Terminated
2004-09-01
Distribution
Nationwide
Recall number
Z-0884-04
Recall status
Terminated
CDRH record ID
34757
FDA classified
2004-08-28