Open original FDA record ↗Recall record
- FDA identifier
- Z-0884-2013
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Feb 26, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.
Source facts
- Action
- Philips Medical Systems issued on October 25, 2012 an Urgent Field Safety Notice for the Digital Diagnost System informing consignees of the software upgrade. The UFSN described the problem, the hazard involved, the action to be taken by the consumer/user and the action planned by Philips. For questions contact your local Philips representative: Technical Support Line 1-800-722-9377.
- Initiated
- 2012-10-25
- Root cause
- Software design
- Terminated
- 2014-03-31
- Distribution
- Worldwide Distribution - USA Nationwide and the Foreign countries of: Austria, Canada, Belgium, China,Denmark, France, germany, Italy, Netherlands,New Zealand, Norway, Philippines, Portugual, SOuth Korea, Spain, Switzerland, Thailand, and UK.
- Recall number
- Z-0884-2013
- Recall status
- Terminated
- CDRH record ID
- 114464
- FDA classified
- 2013-02-26
- Recall event number