A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.
Source facts
Action
The firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.
Initiated
2004-01-05
Root cause
Other
Terminated
2004-09-10
Distribution
Units were distributed to end users worldwide including 67 domestic consignees.
Recall number
Z-0885-04
Recall status
Terminated
CDRH record ID
31967
FDA classified
2004-07-20
Recall event number
28472