Skip to content

Public FDA record

Z-0885-04: EZ-DOP, diagnostic ultrasound imaging system

Recall record Z-0885-04 for DWL Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0885-04
Company in FDA record
DWL Systems, Inc.
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.

Source facts

Action
The firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.
Initiated
2004-01-05
Root cause
Other
Terminated
2004-09-10
Distribution
Units were distributed to end users worldwide including 67 domestic consignees.
Recall number
Z-0885-04
Recall status
Terminated
CDRH record ID
31967
FDA classified
2004-07-20
Recall event number
28472

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗