Skip to content

Public FDA record

Z-0885-2022: SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x

Recall record Z-0885-2022 for Volcano Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0885-2022
Company in FDA record
Volcano Corporation
FDA record creation date
Apr 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.

Source facts

Action
On 01/19/22, correction notices were sent to customers who were informed of the following: If FFR measurement(s) are made prior to an iFR/FFR co-registration in the same procedural session. The user will be visually alerted with a warning message "Insufficient data, distal segment is not co-registered" on the display about the distal segment not being co-registered. The user may also notice the initial iFR/FFR co-registered distal value(s) will be higher than the correctly displayed distal iFR/FFR value(s). Recommended Course of Action: If prior to the iFR/FFR Pullback on the system, FFR measurement(s) were performed within the same procedural session, the user needs to follow any one of the following steps just prior to the iFR/FFR pullback to mitigate the issue: - Click on the "iFR Spot" button on the IntraSight system and perform at least one iFR Spot measurement. OR: - Go back to the "case menu" on the system and then re-enter the LIVE screen to perform iFR pullback. OR: - Exit the procedure and then re-enter the procedure using "Continue Procedure" option. Philips is in the process of updating the Operators Manual to ensure the issue, mitigations, and workarounds are effectively highlighted to inform the user. Notify all system users within your facility of this communication and retain a copy for reference. Send a copy of this notification to any customer to whom you have distributed the affected product to. Customers requiring additional information can contact customer service: 1-800-228-4728, Option 2, igtd.remotesupport@philips.com Hours of Operation: Monday - Friday 8:00AM - 5:00PM PST
Initiated
2022-01-19
Root cause
Device Design
Distribution
US: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, CY, CA, TW, LV, SE, HR, SI, MY, KR, IL, TH, AU, BR, CL, SK, CH, LB, OM, ID, PK, IE, NZ, PR, RO, MM, AR, RS, RU, PA, KE, UZ
Recall number
Z-0885-2022
Recall status
Open, Classified
CDRH record ID
192180
FDA record created
2022-04-06
Recall event number
89673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗